This software updates the firmware of the following camera models to version “3.1.18980”:
The camera’s model designation and type information is given on its type-plate Other than the above listed camera models are not supported by this version of the software.
Firmware “3.1.18980” corrects the following issues:
Cameras within the following serial number ranges are affected:
The serial number of the camera can be found on its type-plate and on the barcode label that is attached to the packaging box.
The current firmware version of the camera can also be verified inside the OLYMPUS application software.
To do so please start your OLYMPUS application (e.g. cellSens, OLYMPUS Stream, PRECiV, DP2-TWAIN) and check the information given under “Device settings” → “Camera” section → “Information”. To access the “Device settings”, please open the “Acquire” menu (PRECiV: “Gear” button in the program’s title bar) → “Devices” → “Device Settings…”.
Cameras with a firmware version lower than “3.1.18980” need to be updated.
Before installation
Download
Updating the firmware
Do NOT disconnect the camera or switch off your computer while the firmware is being updated.
Do NOT shut down or force quit the firmware updater software while the firmware update is running.
Do NOT start/run any application software other than the firmware updater while the update is in progress.
Not following the instructions mentioned above might permanently render the camera inoperable.
Should an error message occur during the firmware update, do NOT disconnect the camera, but repeat the update process for this camera till the update was performed successfully. Close the firmware updater software and restart it before repeating the update process. Click the [Yes] button in step 8. if the message is shown for the camera which failed the firmware update. Do not repeat the update for other connected cameras that were already updated successfully.