2015年 2月 09日 [Important] To Users of the SZ-FLR ESD Fluorescent Ring Illuminator for Stereo Zoom Microscopes Dear Valued OLYMPUS Customer: We, OLYMPUS CORPORATION, have recently discovered that the SZ-FLR ESD fluorescent ring illuminator for stereo zoom microscopes may pose a very small risk of causing an electric shock if not grounded properly. The affected products are listed below. To date there have been no report of electric shocks or product failures. Nonetheless, to ensure continuing safety and reliability, we will inspect the affected products and carry out repairs free of charge. Until the inspection process can be completed, if you have any of the affected products please recheck that they are properly grounded. If you have any of the affected products, please contact the following. People's Republic of China
OLYMPUS (China) Co.,LTD.Shanghai Branch
Japan Customer support center Other Asia,Oceania Please contact your nearest sales representative for Olympus microscopes. The following information should be included: (1) Your Name; (2) Your Country; (3) Product name; (4) Quantity; and (5) Serial numbers (for each product). Or please fill the attached "Application for free inspection of SZ-FLR ESD" sheet and send it. Maintaining the highest quality standards is our highest priority. We assure you that the OLYMPUS group of companies takes this situation very seriously and is taking all possible quality control measures to prevent any recurrence. Thank you for your continuing support of OLYMPUS products. OLYMPUS CORPORATION <Affected Products>Fluorescent ring illuminator for stereo zoom microscopes (Electro-Static Discharge)
SZ-FLR100ESD
The following products are not affected.
<Cause and Description of Fault>Current leakage may occur between the power supply unit and the lamp housing. If the product is used without grounding, there is a risk of electrical shock. If the product is correctly grounded, the leakage breaker in the building may be triggered, but there is no risk of shock.
<Appearance of SZ-FLR ESD, Location of Nameplate, Information in Nameplate>
Nameplates of affected products
Nameplate of unaffected products