Evident LogoOlympus Logo
Thickness and Flaw Inspection Solutions

Specifications

Coil Drive Signals
Excitation Modes Single or multifrequency mode.
Multiplexed mode allowing time multiplexing of up to two simultaneous frequencies.
Number of Coil Drivers 2 independent generator outputs
Number of Input/Time Slots Max 2 inputs per smart probe connector
Max 8 inputs for legacy connector
8 inputs total for the instrument, 32 time slots each for a total of 256 time slots
Frequency Range 50 kHz to 2 MHz
Output Impedance, per ISO 15548-1 Section 6.2.3 < 1.5 Ω typical at 480 kHz, current limited
Maximum Output Load, per ISO 15548-1 Section 6.2.5 Up to 9W typical at 480 kHz
Probe inputs
Input Impedance, 16-pin Connector, per ISO 15548-1 Section 6.3.2 850 kΩ typical
Input Capacitance, 16-pin Connector 250 pF typical
Input Impedance, 41-pin Connector 1 MΩ typical
Input Capacitance, 41-pin Connector 220 pF typical
Absolute Maximum Allowable Input Voltage, per ISO 15548-1 Section 6.3.1 ±10 V (20 V p-p)
Acquisition Inputs
Gain Adjustment range: −26 dB to 76 dB
A/D Converter 24-bit resolution
Digital Filters Programmable digital low-pass and high-pass filters
Maximum Acquisition Rate 40 kHz
Standard
Standard DIN EN ISO 15548
Network Interface 1000BASE-T x-coded 1 Gb
Size (W × H × D) 295 mm x 133 mm x 473 mm
Weight 8.7 kg (19.5 lb)
IP Rating IP55
Total Peak Power Consumption 85 W
DC Input 24 VDC
Temperature Range Operating 0 °C–45 °C (32 °F–113 °F),
Storage −20 °C–70 °C (−4 °F–158 °F)
Digital I/O & Encoders 8 digital inputs (programmable)
8 digital outputs (programmable) 2 encoder inputs (incremental, A & B phase)