Coil Drive Signals | |
---|---|
Excitation Modes |
Single or multifrequency mode.
Multiplexed mode allowing time multiplexing of up to two simultaneous frequencies. |
Number of Coil Drivers | 2 independent generator outputs |
Number of Input/Time Slots |
Max 2 inputs per smart probe connector
Max 8 inputs for legacy connector 8 inputs total for the instrument, 32 time slots each for a total of 256 time slots |
Frequency Range | 50 kHz to 2 MHz |
Output Impedance, per ISO 15548-1 Section 6.2.3 | < 1.5 Ω typical at 480 kHz, current limited |
Maximum Output Load, per ISO 15548-1 Section 6.2.5 | Up to 9W typical at 480 kHz |
Probe inputs | |
Input Impedance, 16-pin Connector, per ISO 15548-1 Section 6.3.2 | 850 kΩ typical |
Input Capacitance, 16-pin Connector | 250 pF typical |
Input Impedance, 41-pin Connector | 1 MΩ typical |
Input Capacitance, 41-pin Connector | 220 pF typical |
Absolute Maximum Allowable Input Voltage, per ISO 15548-1 Section 6.3.1 | ±10 V (20 V p-p) |
Acquisition Inputs | |
Gain | Adjustment range: −26 dB to 76 dB |
A/D Converter | 24-bit resolution |
Digital Filters | Programmable digital low-pass and high-pass filters |
Maximum Acquisition Rate | 40 kHz |
Standard | |
Standard | DIN EN ISO 15548 |
Network Interface | 1000BASE-T x-coded 1 Gb |
Size (W × H × D) | 295 mm x 133 mm x 473 mm |
Weight | 8.7 kg (19.5 lb) |
IP Rating | IP55 |
Total Peak Power Consumption | 85 W |
DC Input | 24 VDC |
Temperature Range |
Operating 0 °C–45 °C (32 °F–113 °F),
Storage −20 °C–70 °C (−4 °F–158 °F) |
Digital I/O & Encoders |
8 digital inputs (programmable)
8 digital outputs (programmable) 2 encoder inputs (incremental, A & B phase) |